Cleaning guide · Tubular modules for ultrafiltration
Cleaning, shut-down & storage T-CUT UF modules in MBR
Applies to all T-CUT UF tubular modules with FRP housing in crossflow MBR systems. Module-specific limits (pH, temperature, pressure, chlorine tolerance) are given in the datasheet of your article – they always take priority.
Cleaning chemicals are corrosive and can react with the process medium. Only trained staff may carry out cleanings.
Corrosive chemicals – follow the safety data sheets of the products
Wear eye protection or a face shield
Wear chemical-resistant gloves and clothing
Depressurize and drain before opening any coupling
Never mix hypochlorite (NaOCl) with acid – toxic chlorine gas is released. Always flush the system with water to neutral pH between an alkaline and an acid stage.
Flush the process medium out first. Cleaning solution meeting process medium can react unexpectedly (e.g. heat generation) and damage membranes, modules and piping.
1When to clean
Clean as needed, not more often – every chemical cleaning stresses the membrane.
Permeate flux drops noticeably at the same TMP and temperature (compare flux normalized to 20 °C).
Pressure drop across the modules rises at the same crossflow.
Scheduled interval defined for your application is reached.
Before a shut-down of more than 24 hours or before preservation (see section 6).
Fouling
A gradual coating of the membrane surface or pores – organic, biological or mineral (scaling). Removed by chemical cleaning.
Clogging
Solids, hair or fibres block the tube inlets – usually a pre-treatment problem. Check each tube by flushing it with water; if blocked, contact CUT.
Never try to unclog tubes mechanically (rods, wires, brushes) – it damages the membrane surface and voids the warranty.
Reference values: record the clean-water flux (CWF) at start-up and before and after the first chemical cleaning. Every later cleaning is judged against these values.
2Cleaning solutions and rules
Solution
Against
Typical recipe
A
Alkaline + oxidant
Organic and biological fouling, silica
NaOH to pH 10–11, plus NaOCl to 250 ppm free chlorine (max. 500 ppm)
B
Acid
Scaling, inorganic fouling
Citric acid to pH 2–3 (typically 2–3 wt%); alternatively nitric or phosphoric acid
C
Enzymatic
Proteins, stubborn organic deposits
Commercial enzymatic cleaner – dosage and temperature as specified by its supplier
D
Preservation
Biofouling during long stops
Sodium bisulfite 1.5 %
Special membrane cleaning products may be used if they stay within the pH and temperature limits of the datasheet. Ask CUT before using any other chemical.
Rules for every cleaning
Use good waterRO permeate, DI water or tap water softened to < 80 ppm CaCO₃ – for the solutions and for flushing.
Always under crossflowRecommended crossflow velocity about 2 m/s (with frequency converter), otherwise the same as in filtration. Permeate line open, permeate flow in the range of normal operation.
Heat slowlyRaise the temperature to about 40 °C within 30–60 min (≈ 1 °C/min). Cool down just as slowly. Never exceed the cleaning temperature in the datasheet (e.g. 60 °C).
Stay within the pH windowCleaning pH 2–11 (see datasheet). Check pH, free chlorine and temperature regularly with meter or test strips and re-dose if needed.
Watch the chlorine budgetTotal chlorine exposure of the membrane is limited (e.g. 250,000 ppm·h). Example: a 2-hour cleaning at 250 ppm uses 500 ppm·h. Keep a log.
Replace dirty solutionIf the solution turns strongly discoloured, drain it, flush, prepare a fresh solution and repeat the stage.
3Routine: flush and backwash
FlushRinse the process medium out with water via the concentrate outlet.
Water circulationFill the tank with DI water, return permeate and concentrate to the tank, circulate for about 20 min.
CheckMeasure the clean-water flux. If it is clearly below the reference, continue with a standard cleaning (section 4).
Optional: backwash during filtration
Short backwash pulses (dark blue) push permeate or clean water back through the membrane and loosen pore blocking.
Every 20–30 min for about 30 s, clean water or permeate is pumped in from the permeate side.
The feed pump runs at lowest speed to carry loosened solids away – it must not build up pressure.
Check that backwash suits your feed before you implement it.
Max. 0.6 bar from permeate to feed side during backwash (e.g. feed 0.2 bar, permeate 0.8 bar) and never above the max. permeate pressure in the datasheet. Higher pressure damages the membrane.
4Standard cleaning
The regular cleaning – in wastewater usually with solution A. Use B (acid) for scaling or C (enzymatic) if A is not effective.
1Flush with water
2Circulate A/B/C30–120 min
3Soak A/B/C30–60 min
4Flush ~20 % of the solutionoptional
5Circulate A/B/C30–120 min
6Flush with waterto neutral pH
FlushRinse the process medium out of the whole system with water.
CirculatePrepare the solution, start circulation under crossflow, heat slowly. Watch flux and pressure drop – the first comparison with the reference is possible here.
SoakStop the pump and let the solution act.
Optional partial flushDrain about 20 % of the solution if a lot of fouling was released during soaking.
Circulate againUntil flux and pressure drop no longer improve.
Flush and verifyFlush with water until pH is neutral, cool down slowly, then measure the clean-water flux and compare with the reference.
Times are guide values – they depend on the installation, the chemicals and the degree of fouling.
5Intensive cleaning
For stubborn fouling, if a standard cleaning is not sufficient. Alternating high and low pH removes deeper deposits. Each stage follows the standard sequence (circulate – soak – optional partial flush – circulate).
Programme
Sequence
Intensive
Flush → A → flush → B → flush → (optional A → flush)
Intensive enzymatic
Flush → C → flush → B → flush → (optional C → flush)
Between A and B always flush to neutral pH. Monitor temperature during heating and cooling – thermal shocks cause irreversible damage.
6Stops, preservation and restart
Once wet, never let the membrane dry out – it can be damaged irreparably.
Duration of the stop
What to do
Up to 24 hours
Flush the modules with water or permeate and keep them filled.
Up to 3 days
Standard cleaning, then fill the system with fresh water.
More than 3 days
Standard or intensive cleaning, then preservation with solution D (sodium bisulfite 1.5 %). Renew the preservation solution at least every 2 months.
Preservation sequence
1Flush with waterskip if just cleaned
2Circulate D30–120 min, heat slowly
3Stop with D insidemax. 2 months
Modules and piping stay filled with the preservation solution. If it turns strongly discoloured, replace it.
Restart after preservation: flush out solution D, run a standard cleaning, then start filtration as described in the installation guide (first start-up).
7Removal, storage and return
Clean before removalRun a standard cleaning so that the feed and permeate side are free of process medium.
Protect against frostIf there is a risk of freezing during storage or transport, soak the cleaned module for several hours in 30 % (to −9 °C) or 40 % (to −15 °C) glycerine solution, then drain completely.
Drain and closeDrain all ports well and close them with end caps or film. Remember: a wet module is much heavier.
Store or packStore dark and frost-free. For transport, pack the module like the original packaging to avoid mechanical damage.
Return to CUTContact CUT before shipping. Returned modules need a completed declaration of decontamination stating the process media and how the module was cleaned.
Heavy load: lift with two people or a hoist and never carry a module by its connections – see weights in the installation guide.
✓Do and don't
Do
Flush out the process medium before every cleaning
Clean under crossflow with the permeate line open
Heat and cool slowly (≈ 1 °C/min)
Check pH, chlorine and temperature during cleaning
Record clean-water flux before and after cleaning
Keep the membrane wet at all times
Don't
Mix hypochlorite with acid
Exceed the pH, temperature or chlorine limits in the datasheet
Unclog tubes mechanically
Backwash above 0.6 bar
Use silicone, ketones, esters, ethers or polar organic solvents